Simple Plex Human C-Reactive Protein/CRP Cartridge

Catalog #: SPCKB-PS-000200 Datasheet / COA / SDS
Simple Plex Human CRP assay kit contains cartridge, sample diluent SD13, and wash buffer. Part # starting with ST01 ships in 2-3 days. Sample type indicated with PS (Plasma/Serum)
Catalog # Availability Size / Price Qty
SPCKB-PS-000200
SPCKA-PS-000200
SPCKC-PS-000200
ST01B-PS-000200
Control Products Available
Human CRP Ella Assay
3 Images
Product Details
FAQs
Supplemental Products
Reviews

Simple Plex Human C-Reactive Protein/CRP Cartridge Summary

Assay Type
Automated sandwich ELISA
Format
Microfluidic cartridge containing three Glass Nano Reactors (GNR)
Assay Length
90 minutes
Sample Type & Volume Required Per Well*
Cell Culture Supernates (10 uL), Serum (10 uL), EDTA Plasma (10 uL), Heparin Plasma (10 uL), Cerebrospinal Fluid (10 uL)
Sensitivity
1.24 pg/mL
Assay Range
32.8 - 50,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Cerebrospinal Fluid)
Specificity
Natural and recombinant human CRP.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.

*The Sample Volume represented is based on the amount of sample incorporated into the reaction after taking into account the assay’s minimum required dilution for a given matrix. Serial dilution may be necessary to achieve some of the final sample volumes represented.

Product Summary

Simple Plex assay for the detection of human C-Reactive Protein (CRP) in cell culture supernatant (CCS), serum and plasma (EDTA/Heparin), and cerebrospinal fluid (CSF).

Precision

Intra-Assay Precision (Precision within an assay) Each control was tested 16 times in one assay
Inter-Assay Precision (Precision between assays) Replicates of each control were tested in multiple assays performed by at least three technicians using two lots of reagents

Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Cerebrospinal Fluid

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 1 2
n 16 16 23 22
Mean (pg/mL) 291 18310 282 18318
Standard Deviation 19.6 783 14.2 1170
CV% 6.7 4.3 5 6.4

Recovery

Recovery at three different spiked concentrations within the range of the assay was evaluated.

Sample Type Average % Recovery Range %
Cell Culture Supernates (n=3) 99 97-101

Linearity

Samples containing and/or spiked with high concentrations of human CRP were serially diluted with Sample Diluent to produce samples within the dynamic range of the assay.
Human CRP Simple Plex Linearity

Scientific Data

Human CRP Ella Standard Curve

Product Datasheets

You must select a language.

x

Preparation and Storage

Shipping
The product is shipped at ambient temperature. Upon receipt, store it immediately at the temperature recommended below.
Storage
Store the unopened product at 2 - 8 °C. Do not use past expiration date.

Background: C-Reactive Protein/CRP

C-Reactive Protein (CRP), also known as Pentraxin 1, is a secreted pentameric protein that functions as a sensor and activator for the innate immune response. In humans, it is a major acute-phase protein; its circulating concentration is dramatically elevated at the onset of inflammation. In mice, however, serum CRP levels increase only slightly during inflammation, and the analogous acute phase role is filled by Pentraxin 2. CRP binds, opsonizes, and induces the phagocytosis of bacteria and apoptotic cells. It regulates activation of the classical complement pathway by binding several proteins in the complement cascade as well as Fc gamma RI, Fc gamma RIIA, and Fc gamma RIIB on macrophages and dendritic cells. It also promotes dendritic cell maturation and humoral immunity. In cardiovascular disease, CRP binds to oxidized LDL, exacerbates tissue damage in myocardial infarction, and inhibits the repair of injured vascular endothelium.

Entrez Gene IDs:
1401 (Human); 12944 (Mouse); 25419 (Rat)
Alternate Names:
C-Reactive Protein; C-reactive protein, pentraxin-related; CRP; MGC88244; pentraxin 1; PTX1MGC149895

FAQs

No product specific FAQs exist for this product, however you may

View all Simple Plex Assays FAQs
Loading...

Reviews for Simple Plex Human C-Reactive Protein/CRP Cartridge

There are currently no reviews for this product. Be the first to review Simple Plex Human C-Reactive Protein/CRP Cartridge and earn rewards!

Have you used Simple Plex Human C-Reactive Protein/CRP Cartridge?

Submit a review and receive an Amazon gift card.

$25/€18/£15/$25CAN/¥75 Yuan/¥2500 Yen for a review with an image

$10/€7/£6/$10 CAD/¥70 Yuan/¥1110 Yen for a review without an image

Submit a Review