Human XL Cytokine Luminex® Kit Performance Assay

Build your own Luminex Assay with our Luminex Assay Customization Tool.

U.S. orders usually ship in 5-10 business days

Build Now
Human Luminex Cytokine Assay Standard Curves
2 Images
Product Details
Citations (6)
FAQs
Supplemental Products
Reviews (1)

Human XL Cytokine Luminex® Kit Performance Assay Summary

Panel Luminex Performance Human XL Cytokine Panel
Assay Type Magnetic bead-based multiplex assay for the Luminex® platform
Format 1 x 96-well microplate and Magnetic antibody-coated beads
Species Human
Analytes Detected Please see analyte list in assay customization tool below.
Performance Validation Our Luminex High Performance Assays undergo our most extensive validation testing. All Luminex High Performance Assay Panels are validated for use with serum and plasma samples. Some High Performance Assays are additionally validated for cell culture supernatants, milk, saliva, or urine, as indicated on the individual product data sheets. Luminex High Performance Assays are tested for sensitivity (three-quarters the low standard), intra-assay precision, inter-assay precision, and to ensure assay linearity for validated sample types. Recovery values for individual samples in validated sample types are also tested. Validation data for each analyte can be found in the product datasheet.

Make your selections below to configure your assay

Filter Analytes
Select All Select None

Assays

Below are kits that are available based on your selections. These selections may be further customized before adding to the cart or emailing the configuration to your purchasing department or local distributor to place your order. Selected analytes may be included in more than one High Performance Assay. All assay options are displayed.

If you run into bead region incompatibilities, send your Luminex code to our team and find out more.

How would you like these analytes mixed?

Premixed
User Mixed
Please select at least one mixing option to see the kits.

Would you like to split based on dilution?

Do not split
Split based on serum
Split based on plasma
Split based on cell culture supernates

Not finding your analyte of interest? View our complete Luminex offering in the full version of our Luminex Assay Customization Tool. We also offer Custom Luminex Services.

Product Features

Magnetic Luminex High Performance Assays are flexible bead-based multiplex assays. They allow up to 50 user-defined target analytes to be simultaneously profiled using cell culture supernates, serum, or plasma samples. Some panels are also validated with human milk, saliva, and/or urine samples.

  • Simultaneously profile up to analytes of your choice in one sample
  • Magnetic format allows for easier wash steps
  • Undergoes similar validation testing as Quantikine® ELISA Kits
  • Precise and Specific
  • Requires a small sample volume (<50 µL)
  • Run your assay in just 3-5 hours
  • Mass-calibrated standards for consistent results with every new lot of material
  • Enough reagents for one 96-well plate
Kit Components
  • Premixed cocktail of antibody-coated Magnetic beads
  • Premixed cocktail of biotinylated detection antibody
  • Standard Cocktail(s)
  • Bead Diluent
  • Biotin Antibody Diluent
  • Calibrator Diluent
  • Wash Buffer
  • Streptavidin-PE
  • One flat-bottom 96-well Microplate
  • Foil Plate Sealers (4)
  • Mixing Bottle
  • Standard Value Card**
  • High and low controls with Control Mean Value Card

**A standard curve must be generated each time an assay is run, utilizing values from the Standard Value Card included in the Kit.

Precision

Intra-Assay Precision - is generated from the mean of the %CV's from 40 reportable results across two different concentrations of analytes in a single serum assay

Inter-Assay Precision - is generated from the mean of the %CV's across two different concentrations of analytes across 30 different serum assays.

  n=40 (20x2 levels) n=30
Analyte Intra-Assay (%CV) Inter-Assay (%CV)
CD40 Ligand 3.8 10.8
EGF 4.1 8.0
Eotaxin 4.8 7.4
FGF basic 2.0 7.6
Flt-3 Ligand 2.4 8.4
G-CSF 1.8 7.0
GM-CSF 3.4 11.2
Granzyme B 2.9 8.2
GRO alpha 1.9 5.6
GRO beta 1.9 6.8
IFN-alpha 2 3.6 10.0
IFN-beta 2.4 7.1
IFN-gamma 2.3 10.9
IL-1 alpha 5.0 10.6
IL-1 beta 2.8 10.1
IL-1ra 3.6 12.4
IL-2 3.9 9.6
IL-3 3.6 9.9
IL-4 2.4 8.4
IL-5 1.8 8.3
IL-6 2.2 8.8
IL-7 2.8 9.6
IL-8 3.2 84
IL-9 6.2 10.3
IL-10 3.6 11.2
IL-12 p70 2.6 9.8
IL-13 2.5 8.3
IL-15 3.2 11.2
IL-17A 2.5 9.8
IL-17E 3.1 8.1
IL-33 2.9 9.7
IP-10 3.3 6.6
MCP-1 1.8 6.3
MIP-1 alpha 2.9 5.7
MIP-1 beta 1.7 7.9
MIP-3 alpha 3.0 9.4
MIP-3 beta 2.5 8.6
PDGF-AA 2.7 7.1
PDGF-AB/BB 2.2 7.4
PD-L1/B7-H1 2.7 8.3
RANTES 1.8 7.5
TGF-alpha 4.4 9.2
TNF-alpha 3.0 10.3
TNF-beta 2.2 8.0
TRAIL 2.4 7.9
VEGF 2.9 10.1

Accuracy

Linearity - The data represents mean spiked or natural linearity in serum matrix samples. Samples with natural linearity are marked with an asterisk.

Spike Recovery - The data represents mean percent recovery of spiked standards ranging from low, medium, and high concentration in serum matrix samples.

  Linearity Spiked
Recovery
Analyte N Average % N Average %

*Natural linearity
HE = Spiked recovery was not observed due to natural linearity in high endogenous samples

CD40 Ligand 4 131 4 75
EGF* 4 150 HE
Eotaxin 4 68 4 156
FGF basic 4 128 4 87
Flt-3 Ligand 4 119 4 78
G-CSF 4 208 4 34
GM-CSF 4 101 4 98
Granzyme B 4 107 4 96
GRO alpha 4 97 4 139
GRO beta* 4 88 HE
IFN-alpha 2 4 106 4 87
IFN-beta 4 137 4 63
IFN-gamma 4 127 4 73
IL-1 alpha 4 143 3 74
IL-1 beta 4 106 4 94
IL-1ra* 4 126 HE
IL-2 4 115 4 88
IL-3 4 123 4 78
IL-4 4 141 4 58
IL-5 4 113 4 88
IL-6 4 126 4 78
IL-7 4 110 4 90
IL-8 4 118 4 78
IL-9 4 111 4 96
IL-10 4 117 4 90
IL-12 p70 4 111 4 87
IL-13 4 118 4 89
IL-15 4 103 4 76
IL-17A 4 128 4 72
IL-17E 4 150 4 55
IL-33 4 103 4 87
IP-10 4 124 4 72
MCP-1* 4 80 HE
MIP-1 alpha 4 95 4 111
MIP-1 beta 4 109 4 95
MIP-3 alpha 4 115 4 99
MIP-3 beta 4 123 4 76
PDGF-AA* 4 91 HE
PDGF-AB/BB* 4 108 HE
PD-L1/B7-H1 4 106 4 96
RANTES* 4 106 HE
TGF-alpha 4 142 4 80
TNF-alpha 4 134 4 71
TNF-beta 4 131 4 65
TRAIL 4 144 4 69
VEGF* 4 129 HE
Which Luminex Assay is right for me?

Assays for the Luminex platform are offered as High Performance Assays or Assays. The Luminex High Performance Assays are fully validated panels with a focused selection of analytes. They are available in We Mix, You Mix, and Predetermined formats. The Luminex Assays allows for the maximum number of analytes in a multiplex and is supplied as a premixed kit. View a table comparing the features and benefits of these bead-based multiplex assays.

More Information About Magnetic Bead-based Multiplex Kits For The Luminex® Platform

Magnetic Luminex High Performance Assays are designed for use with the Luminex MAGPIX CCD Imager. Alternatively, kits can be used with the Luminex 100/200™ or FLEXMAP 3D, dual laser, flow-based sorting and detection platforms.

Analyte-specific antibodies are pre-coated onto color-coded microparticles. Microparticles, standards, and samples are pipetted into wells and the immobilized antibodies bind the analytes of interest. After washing away any unbound substances, a biotinylated antibody cocktail specific to the analytes of interest is added to each well. Following a wash to remove any unbound biotinylated antibody, Streptavidin-Phycoerythrin conjugate (Streptavidin-PE), which binds to the biotinylated detection antibodies, is added to each well. A final wash removes unbound Streptavidin-PE and the microparticles are resuspended in buffer and read using the Luminex or Bio-Plex analyzer. One laser is bead-specific and determines which analyte is being detected. A magnet in the analyzer captures and holds the superparaMagnetic microparticles in a monolayer. Two spectrally distinct Light Emitting Diodes (LEDs) illuminate the beads. One LED identifies the analyte that is being detected and the second LED determines the magnitude of the PE-derived signal, which is in direct proportion to the amount of analyte bound. Each well is imaged with a CCD camera. Kits can also be used with Luminex 100/200 or a Bio-Rad Bio-Plex dual laser, flow-based systems.

Scientific Data

Human Luminex Cytokine Assay Standard Curves

Human Luminex Cytokine Assay Serum Sample Values Serum samples from 20 apparently healthy individuals were run on the Human XL Cytokine Luminex High Performance Panel and sample values plotted. Minimum and maximum values shown by exterior lines, the box represents the 1st and 3rd quartiles with the center line representing the median.

Product Datasheets

You must select a language.

x

Citations for Human XL Cytokine Luminex® Kit Performance Assay

R&D Systems personnel manually curate a database that contains references using R&D Systems products. The data collected includes not only links to publications in PubMed, but also provides information about sample types, species, and experimental conditions.

6 Citations: Showing 1 - 6
Filter your results:

Filter by:

  1. Differences in Eotaxin Serum Levels between Polytraumatized Patients with and without Concomitant Traumatic Brain Injury-A Matched Pair Analysis
    Authors: Negrin, LL;Ristl, R;Wollner, G;Hajdu, S;
    Journal of clinical medicine
    Species: Human
    Sample Types: Serum
  2. Helminth exposure and immune response to the two-dose heterologous Ad26.ZEBOV, MVA-BN-Filo Ebola vaccine regimen
    Authors: Barry, H;Lhomme, E;Surénaud, M;Nouctara, M;Robinson, C;Bockstal, V;Valea, I;Somda, S;Tinto, H;Meda, N;Greenwood, B;Thiébaut, R;Lacabaratz, C;
    PLoS neglected tropical diseases
    Species: Human
    Sample Types: Serum
  3. Protocol for a multicentre cross-sectional, longitudinal ambulatory clinical trial in rheumatoid arthritis and Parkinson's disease patients analysing the relation between the gut microbiome, fasting and immune status in Germany (ExpoBiome)
    Authors: Hansen, B;Laczny, CC;Aho, VTE;Frachet-Bour, A;Habier, J;Ostaszewski, M;Michalsen, A;Hanslian, E;Koppold, DA;Hartmann, AM;Steckhan, N;Mollenhauer, B;Schade, S;Roomp, K;Schneider, JG;Wilmes, P;
    BMJ open
    Species: Human
    Sample Types: Plasma
  4. Dissecting human population variation in single-cell responses to SARS-CoV-2
    Authors: Aquino, Y;Bisiaux, A;Li, Z;O'Neill, M;Mendoza-Revilla, J;Merkling, SH;Kerner, G;Hasan, M;Libri, V;Bondet, V;Smith, N;de Cevins, C;Ménager, M;Luca, F;Pique-Regi, R;Barba-Spaeth, G;Pietropaoli, S;Schwartz, O;Leroux-Roels, G;Lee, CK;Leung, K;Wu, JT;Peiris, M;Bruzzone, R;Abel, L;Casanova, JL;Valkenburg, SA;Duffy, D;Patin, E;Rotival, M;Quintana-Murci, L;
    Nature
    Species:  Human
    Sample Types: Cell Culture Supernates
  5. First-line durvalumab and tremelimumab with chemotherapy in RAS-mutated metastatic colorectal cancer: a phase 1b/2 trial
    Authors: Thibaudin, M;Fumet, JD;Chibaudel, B;Bennouna, J;Borg, C;Martin-Babau, J;Cohen, R;Fonck, M;Taieb, J;Limagne, E;Blanc, J;Ballot, E;Hampe, L;Bon, M;Daumoine, S;Peroz, M;Mananet, H;Derangère, V;Boidot, R;Michaud, HA;Laheurte, C;Adotevi, O;Bertaut, A;Truntzer, C;Ghiringhelli, F;
    Nature medicine
    Species: Human
    Sample Types: Plasma
  6. Peripheral immune cell death in sepsis based on bulk RNA and single-cell RNA sequencing
    Authors: Liu, S;Duan, C;Xie, J;Zhang, J;Luo, X;Wang, Q;Liang, X;Zhao, X;Zhuang, R;Zhao, W;Yin, W;
    Heliyon
    Species: Human
    Sample Types: Plasma

FAQs

No product specific FAQs exist for this product, however you may

View all Luminex Assay FAQs
Loading...

Reviews for Human XL Cytokine Luminex® Kit Performance Assay

Average Rating: 4 (Based on 1 Review)

5 Star
0%
4 Star
100%
3 Star
0%
2 Star
0%
1 Star
0%

Have you used Human XL Cytokine Luminex® Kit Performance Assay?

Submit a review and receive an Amazon gift card.

$25/€18/£15/$25CAN/¥75 Yuan/¥2500 Yen for a review with an image

$10/€7/£6/$10 CAD/¥70 Yuan/¥1110 Yen for a review without an image

Submit a Review

Filter by:


Human XL Cytokine Luminex® Performance Panel Premixed Kit
By Lana Jagannathan on 11/03/2023
Sample Tested: Cell culture supernatant

The workflow for the Luminex is incredibly long (8 hours minimum), but takes only one full day to perform. Before performing Luminex, ensure that your machine has been serviced recently and is in working order, since the variability between samples can be influenced by faulty fluidics.